Comparison of NMR and MS
The two most common techniques used in data acquisition are nuclear magnetic resonance and mass spectrometry. Table 1 shows some of the key differences between the two techniques.
Nuclear magnetic resonance (NMR) | Mass spectrometry (MS) | |
Sensitivity | Low | High |
Reproducibility | Very high | Average |
Number of detectable metabolites | 30-100 | 300-1000+ (depending on whether GC-MS or LC-MS is used) |
Targeted analysis | Not optimal for targeted analysis | Better for targeted analysis than NMR |
Sample preparation | Minimal sample preparation required | More complex sample preparation required |
Tissue extraction | Not required – tissues can be analysed directly | Requires tissue extraction |
Sample analysis time | Fast – the entire sample can be analysed in one measurement | Longer than NMR – requires different chromatography techniques depending on the metabolites analysed |
Instrument Cost | More expensive and occupies more space than MS | Cheaper and occupies less space than NMR |
Sample Cost | Low cost per sample | High cost per sample |